Doc# 0296-0018 • REV R (July 2017) Page 52 of 61
Step 5. Turn the TTS Amplitude control knob to the minimum amplitude.
Step 6. Set TTS to operate in the Sensivity Mode, select Displace ment Sensor Type and to read sensi-
vity in Mils. (Test Page A)
Step 7. Press the 100 Hz Frequency Control buon.
Step 8. Press the RUN/STORE buon.
Step 9. Set the Amplitude Control to as indicated in the Sample Data Sheet (from 2 to 5 Mils) and verify
sensivity is within the require range. (See Table 5-4)
Dynamic Sensivity Test
Gap Mils Amplitude Mils SF mV/MIL
Requirement mV/
MIL
10 2 190-210
20 5 190-210
30 5 190-210
40 5 190-210
50 5 190-210
60 5 190-210
70 5 190-210
80 5 190-210
TABLE 5-4: SAMPLE DATA SHEET
Step 10. Turn the Amplitude knob counterclockwise unl the knob stops. Regap the probe to verify
sensivity over the probes specied calibraon range.
Step 11. Repeats steps 9 and 10 unl all data points listed in the Sample Data sheet have been veried.