EasyManua.ls Logo

Mettler Electronics Sonicator 740 - Page 16

Mettler Electronics Sonicator 740
38 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Mettler Electronics Corp. — Rev.G_06/28/12
16
Guidance and manufacturer’s declaration – electromagnetic immunity
The Sonicator 740 is intended for use in the electromagnetic environment specified below. The customer or
the user of the Sonicator 740 should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance
level
Electromagnetic environment - guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 GHz
3 V/m
80 MHz to 2.5 GHz
3 V
3 V/m
Portable and mobile RF communications
equipment should be used no closer to any part
of the Sonicator 740, including cables, than the
recommended separation distance calculated
from the equation applicable to the frequency of
the transmitter.
Recommended separation distance
d = 1.2P
d = 1.2P 80MHz to 800 MHz
d = 2.3P 800MHz to 2.5 GHz
where P is the maximum output power rating of
the transmitter in watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
a
should be less than the compliance level in each
frequency range.
b
Interference may occur in the vicinity of
equipment marked with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios,
amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field
strength in the location in which the Sonicator 740 is used exceeds the applicable RF compliance level above, the Sonicator 740
should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as
reorienting or relocating the Sonicator 740.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.