Mettler Electronics Corp. — Rev.G_06/28/12
26
7.
5
Output power is not within output tolerances. To attempt to clear this error:
unplug the unit from the wall outlet, remove the battery, reinstall the battery
and plug the unit back in. Turn the unit back on and attempt to restart the
unit. If the error is not resolved, the Sonicator 740 requires servicing.
8.
6
Battery is over temperature. Unplug the unit from the wall outlet and
remove or replace the battery.