Micro Direct MicroLab - Calibration Check Report; Calibration Error Interpretation

Default Icon
40 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
16
The calibration error for expiration
and inspiration at each flow rate
are displayed. The calibration
error should be less than 3.5%. If
a greater error is shown, repeat
the procedure ensuring that the
syringe is emptied and filled in a
smooth manner without jerking the
handle. If an error greater than
3.5% is still shown, inspect the
turbine transducer and clean if
necessary.