Power supply wiring terminalsFigure 1-13:
A. Barrier wiring is applicable to intrinsically safe installations only
5. Perform a Known Density Verification (KDV) check.
Use the Known Density Verification procedure to match the current meter
calibration with the factory calibration. If the meter passes the test, then it has not
drifted or changed during shipment.
For more information on performing a KDV check, see the configuration and use
manual that shipped with the product.
Planning
10 Micro Motion Compact Density Meter