2. Set the isolator switch settings for the TPS/DO connection (MTL5532 isolator). You
must set the isolator switches appropriately for Pins 1 through 5 (see Table 3-4).
The switches are located on the side of the isolator, and must be set to either Off (the
up position) or On (the down position).
MTL5532 switch location (plus ON/OFF switch position)Figure 3-6:
MTL5532 switch settingsTable 3-4:
Switch ON/OFF?
1 ON
2 OFF
3 OFF
4 OFF
3.4 Processor wiring for remote-mount 2700
FOUNDATION fieldbus
™
option
3.4.1 RS-485 entity parameters for the remote-mount 2700
FOUNDATION fieldbus
™
option
DANGER!
Hazardous voltage can cause severe injury or death. To reduce the risk of hazardous voltage,
shut off power before wiring the meter.
Wiring
28 Micro Motion Compact Density Meter