aponorm
®
by
18
* Please consult your doctor, if this or any other problem occurs
repeatedly.
If you think the results are unusual, please read through
the information in «Section 1.» carefully.
10. Safety, Care, Accuracy Test and Disposal
Safety and protection
• Thisdevicemayonlybeusedforthepurposesdescribed
in this instruction. The manufacturer cannot be held liable
for damage caused by incorrect application.
• Thisdevicecomprisessensitivecomponentsandmustbe
treated with caution. Observe the storage and operating con
-
ditions described in the «Technical Specifications» section.
• Protectitfrom:
water and moisture
impact and dropping
contamination and dust
direct sunlight
heat and cold
• Thecuffsaresensitiveandmustbehandledwithcare.
• Onlypumpupthecuffoncetted.
• Donotusethisdeviceclosetostrongelectromagnetic
fields such as mobile telephones or radio installations.
• Donotusethisdeviceifyouthinkitisdamagedornotice
anything unusual.
• Neveropenthisdevice.
• Ifthedeviceisnotgoingtobeusedforaprolongedperiod
the batteries should be removed.
• Readtheadditionalsafetyinformationintheindividual
sections of this instructions.
Ensure that children do not use this device unsupervised;
some parts are small enough to be swallowed.
Error
«ERR 3»
17
«ERR 5»
«HI»
«LO»
Description
No pressure
in the cuff
Abnormal
result
Pulse or cuff
pressure too
high
Pulse too
low
Potential cause and remedy
An adequate pressure cannot be gene
-
rated in the cuff. A leak may have
occurred. Check that the cuff is cor
-
rectly connected and is not too loose.
Replacethebatteriesifnecessary.
Repeatthemeasurement.
The measuring signals are inaccurate
and no result can therefore be dis
-
played.Readthroughthechecklistfor
performing reliable measurements and
then repeat the measurement.*
The pressure in the cuff is too high
(over300mmHg)ORthepulseistoo
high (over 200 beats per minute).
Relaxfor5minutesandrepeatthe
measurement.*
The pulse is too low (less than 40
beatsperminute).Repeatthemeasu
-
rement.*