aponorm
®
by
18
* Please consult your doctor, if this or any other problem occurs
repeatedly.
If you think the results are unusual, please read through
the information in «Section 1.» carefully.
10. Safety, Care, Accuracy Test and Disposal
Safety and protection
• Thisdevicemayonlybeusedforthepurposesdescribed
in this instruction. The manufacturer cannot be held liable
for damage caused by incorrect application.
• Thisdevicecomprisessensitivecomponentsandmustbe
treated with caution. Observe the storage and operating con
-
ditions described in the «Technical Specifications» section.
• Protectitfrom:
water and moisture
impact and dropping
contamination and dust
direct sunlight
heat and cold
• Thecuffsaresensitiveandmustbehandledwithcare.
• Onlypumpupthecuffoncetted.
• Donotusethisdeviceclosetostrongelectromagnetic
fields such as mobile telephones or radio installations.
• Donotusethisdeviceifyouthinkitisdamagedornotice
anything unusual.
• Neveropenthisdevice.
• Ifthedeviceisnotgoingtobeusedforaprolongedperiod
the batteries should be removed.
• Readtheadditionalsafetyinformationintheindividual
sections of this instructions.
Ensure that children do not use this device unsupervised;
some parts are small enough to be swallowed.
Error
«ERR 3»
17
«ERR 5»
«HI»
«LO»
Description
No pressure
in the cuff
Abnormal
result
Pulse or cuff
pressure too
high
Pulse too
low
Potential cause and remedy
An adequate pressure cannot be gene
-
rated in the cuff. A leak may have
occurred. Check that the cuff is cor
-
rectly connected and is not too loose.
Replacethebatteriesifnecessary.
Repeatthemeasurement.
The measuring signals are inaccurate
and no result can therefore be dis
-
played.Readthroughthechecklistfor
performing reliable measurements and
then repeat the measurement.*
The pressure in the cuff is too high
(over300mmHg)ORthepulseistoo
high (over 200 beats per minute).
Relaxfor5minutesandrepeatthe
measurement.*
The pulse is too low (less than 40
beatsperminute).Repeatthemeasu
-
rement.*

Other manuals for Microlife Aponorm Basis Control

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Microlife Aponorm Basis Control and is the answer not in the manual?

Microlife Aponorm Basis Control Specifications

General IconGeneral
BrandMicrolife
ModelAponorm Basis Control
CategoryBlood Pressure Monitor
LanguageEnglish

Related product manuals