detected blood pressure value. There will be 15 seconds resting time
in-between each measurement. A count down indicates the remaining
time and a beep will sound 5 seconds before the 2nd and 3rd readings
will begin. In case that the single data of each cycle differ too much
from each other, a fourth measurement is performed before the result
will be displayed. In rare cases the blood pressure is such unstable
that even after four measurements the data vary too much. In this
case «ERR 6» is shown and no result can be given.
If one of measurement causes an error message it is repeated.
5.6. Discontinuing a measurement
If it is necessary to interrupt a blood pressure measurement for any
reason (e.g. the patient feels unwell), the 0/I button can be pressed
at any time.
The device immediately lowers the cuff pressure automatically.
5.7. Memory – recall of measurements
This device automatically stores the last 99 measurements. By
pressing the MEMORY button, the number of stored measurements
is displayed (e.g. 30 set). By pressing again, the last measurement,
e.g. MR30 as well as previous 29 measurements (MR29, MR28,
…MR1) can be displayed one after the other. Data obtained in
MAM mode are indicated by a « » symbol.
30 sets of stored measurements – MR30: Value of the last measurement – MR29: Value of the measurement before MR 30
5.8.Memory– cancellation of all measurements
Attention!
Before you delete all readings stored in the memory, make sure you
will not need refer to the readings at a later date.
In order to delete all stored readings, depress the MEMORY button for
at least 7 seconds, the display will show the symbol «CL» and 3 short
beep sounds will be heard to indicate deletion of stored readings.
Note:
If your last stored measurement was taken in the MAM mode, you
will see the « » symbol approximately 3 seconds during the
deletion procedure. (The display will show this symbol during
deletion of all stored measurements, if the last measurement is
taken in the MAM mode).
12