EasyManuals Logo

Microlife BP A100 Plus User Manual

Microlife BP A100 Plus
27 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #9 background imageLoading...
Page #9 background image
7BP A100 Plus
EN
12. Safety, Care, Accuracy Test and Disposal
Safety and protection
Instrument care
Clean the instrument only with a soft, dry cloth.
Cleaning the cuff
Carefully remove spots on the cuff with a damp cloth and soapsuds.
Accuracy test
We recommend this instrument is tested for accuracy every
2 years or after mechanical impact (e.g. being dropped). Please
contact Microlife-Service to arrange the test (see foreword).
Disposal
13. Guarantee
This instrument is covered by a 5 year guarantee from the date
of purchase. The guarantee is valid only on presentation of the
guarantee card completed by the dealer (see back) confirming date
of purchase or the receipt.
Please contact Microlife-Service (see foreword).
14. Technical Specifications
This device complies with the requirements of the Medical Device
Directive 93/42/EEC. Technical alterations reserved.
This instrument may be used only for the purpose described in
this booklet. The manufacturer cannot be held liable for damage
caused by incorrect application.
This instrument comprises sensitive components and must be
treated with caution. Observe the storage and operating condi-
tions described in the «Technical Specifications» section!
Protect it from:
water and moisture
extreme temperatures
impact and dropping
contamination and dust
direct sunlight
heat and cold
The cuffs are sensitive and must be handled with care.
Only pump up the cuff once fitted.
Do not use the instrument close to strong electromagnetic fields
such as mobile telephones or radio installations.
Do not use the instrument if you think it is damaged or notice
anything unusual.
Never open the instrument.
If the instrument is not going to be used for a prolonged period
the batteries should be removed.
Read the additional safety instructions in the individual sections
of this booklet.
Ensure that children do not use the instrument unsuper-
vised; some parts are small enough to be swallowed.
WARNING: Do not wash the cuff in a washing machine or
dishwasher!
Batteries and electronic instruments must be disposed of in
accordance with the locally applicable regulations, not with
domestic waste.
Batteries, cuff and wearing parts are not included.
Opening or altering the instrument invalidates the guarantee.
The guarantee does not cover damage caused by improper
handling, discharged batteries, accidents or non-compliance
with the operating instructions.
Operating tempera-
ture:
10 - 40 °C / 50 - 104 °F
15 - 95 % relative maximum humidity
Storage temperature:
-20 - +55 °C / -4 - +131 °F
15 - 95 % relative maximum humidity
Weight:
748 g (including batteries)
Dimensions:
160 x 140 x 98 mm
Measuring proce-
dure:
oscillometric, corresponding to Korotkoff
method: Phase I systolic, Phase V diastolic
Measurement range:
20 - 280 mmHg – blood pressure
40 - 200 beats per minute – pulse
Cuff pressure display
range:
0 - 299 mmHg
Resolution:
1 mmHg
Static accuracy:
pressure within ± 3 mmHg
Pulse accuracy:
± 5 % of the readout value
Voltage source:
4
x 1.5 V Batteries; size AA
Mains adapter DC 6V, 600 mA (optional)
Reference to
standards:
EN 1060-1 /-3 /-4; IEC 60601-1;
IEC 60601-1-2 (EMC)

Table of Contents

Other manuals for Microlife BP A100 Plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Microlife BP A100 Plus and is the answer not in the manual?

Microlife BP A100 Plus Specifications

General IconGeneral
TypeAutomatic
Placement supportedUpper arm
Mean arterial pressureYes
Diastolic blood pressure-
Battery typeAA
Number of batteries supported4
Display typeLCD

Related product manuals