7Microlife BP A150 AFIB
EN
9. Using a mains adapter
You can operate this device using the Microlife mains adapter (DC
6V, 600 mA).
1. Plug the adapter cable into the mains adapter socket 8 in the
blood pressure monitor.
2. Plug the adapter plug into the wall socket.
When the mains adapter is connected, no battery current is
consumed.
10.Error messages
If an error occurs during the measurement, the measurement is
interrupted and an error message, e.g. «ERR 3», is displayed.
* Please immediately consult your doctor, if this or any other
problem occurs repeatedly.
11.Safety, care, accuracy test and disposal
Follow instructions for use. This document provides important
product operation and safety information regarding this device.
Please read this document thoroughly before using the device
and keep for future reference.
This device may only be used for the purposes described in
these instructions. The manufacturer cannot be held liable for
damage caused by incorrect application.
This device comprises sensitive components and must be
treated with caution. Observe the storage and operating condi-
tions described in the «Technical specifications» section.
The cuffs are sensitive and must be handled with care.
Only pump up the cuff once fitted.
Do not use this device if you think it is damaged or notice
anything unusual.
Never open this device.
Read the additional safety information provided within the indi-
vidual sections of this instruction manual.
Only use the Microlife mains adapter available as an orig-
inal accessory appropriate for your supply voltage.
Ensure that neither the mains adapter nor the cable are
damaged.
Error Descrip-
tion
Potential cause and remedy
«ERR 1» Signal too
weak
The pulse signals on the cuff are too
weak. Re-position the cuff and repeat
the measurement.*
«ERR 2»
BK
Error signal During the measurement, error signals
were detected by the cuff, caused for
instance by movement or muscle
tension. Repeat the measurement,
keeping your arm still.
«ERR 3»
BL
No pres-
sure in the
cuff
An adequate pressure cannot be gener-
ated in the cuff. A leak may have
occurred. Check that the cuff is correctly
connected and is not too loose. Replace
the batteries if necessary. Repeat the
measurement.
«ERR 5» Abnormal
result
The measuring signals are inaccurate
and no result can therefore be displayed.
Read through the checklist for
performing reliable measurements and
then repeat the measurement.*
«ERR 6» AFIB/MAM
Mode
There were too many errors during the
measurement in AFIB/MAM mode,
making it impossible to obtain a final
result. Read through the checklist for
performing reliable measurements and
then repeat the measurement.*
«HI» Pulse or
cuff pres-
sure too
high
The pressure in the cuff is too high (over
299 mmHg) OR the pulse is too high
(over 200 beats per minute). Relax for 5
minutes and repeat the measurement.*
«LO» Pulse too
low
The pulse is too low (less than 40 beats
per minute). Repeat the measurement.*
If you think the results are unusual, please read through the
information in «Section 1.» carefully.
Safety and protection
Error Descrip-
tion
Potential cause and remedy

Table of Contents

Other manuals for Microlife BP A150 AFIB

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Microlife BP A150 AFIB and is the answer not in the manual?

Microlife BP A150 AFIB Specifications

General IconGeneral
TypeAutomatic
Cuff size22 - 42 cm
Number of users1 user(s)
Memory registers1
Placement supportedUpper arm
Mean arterial pressureYes
Diastolic blood pressure-