4
6. During the measurement, the pulse indicator BM flashes in the
display.
7. The result, comprising the systolic AM and the diastolic AN blood
pressure and the pulse rate AO are displayed. Note also the
explanations on further display symbols in this booklet.
8. When the device has finished measuring, remove the cuff.
9. Switch off the device. (The monitor does switch off automati-
cally after approx. 1 min.).
You can stop the measurement at any time by pressing the
ON/OFF button (e.g. if you feel uneasy or an unpleasant
pressure sensation).
How not to store a reading
As soon as the reading is displayed press and hold the ON/OFF
button 1 until «M» BO is flashing. Confirm to delete the reading by
pressing the MAM button 4.
«CL» is displayed when the reading is deleted from the
memory successfully.
How do I evaluate my blood pressure
The triangle on the left-hand edge of the display BL points at the
range within which the measured blood pressure value lies. The
value is either within the optimum (green), elevated (yellow) or
high (red) range. The classification corresponds to the following
ranges defined by international guidelines (ESH, ESC, JSH). Data
in mmHg.
The higher value is the one that determines the evaluation.
Example: a blood pressure value of 140/80 mmHg or a value of
130/90 mmHg indicates «blood pressure too high».
Appearance of the irregular heartbeat (IHB) symbol
This symbol BT indicates that an irregular heartbeat was detected.
In this case, the measured blood pressure may deviate from your
actual blood pressure values. It is recommended to repeat the
measurement.
4. Data memory
This device automatically stores up to 99 measurement values for
each of the 2 users.
Select either user 1 or 2 by pressing the user button 5.
Viewing the stored values
Press the M-button 3 briefly, when the device is switched off. The
display first shows «M» BO, and «A»which stands for the average
of all stored values.
Pressing the M-button again displays the previous value. Pressing
the M-button repeatedly enables you to move from one stored
value to another.
Blood pressure readings with suboptimal cuff fit AR-A are
not considered in the average value.
Pay attention that the maximum memory capacity of 99
memories is not exceeded. When the 99 memory is full,
the oldest value is automatically overwritten with the
100 value. Values should be evaluated by a doctor before
the memory capacity is reached – otherwise data will be
lost.
Clearing all values
Make sure the correct user is activated.
If you are sure that you want to permanently remove all stored
values, hold down the M-button (the device must have been
switched off beforehand) until «CL ALL» appears and then
release the button. To permanently clear the memory, press the
MAM button while «CL ALL» is flashing. Individual values
cannot be cleared.
Cancel deletion: press ON/OFF button 1 while «CL
ALL» is flashing.
Range Systolic Diastolic Recommendation
1. blood pressure
too high
≥135 ≥85 Seek medical advice
2. blood pressure
elevated
130 - 134 80 - 84 Self-check
3. blood pressure
normal
<130 <80 Self-check
Information for the doctor in case of repeated appearance
of the IHB symbol:
This device is an oscillometric blood pressure monitor that also
measures the pulse during blood pressure measurement and
indicates when the heart rate is irregular.