9. Error messages/Troubleshooting
If an error occurs during a measurement, the measurement is dis-
continued and a corresponding error code is displayed (Example:
Error No. 2).
Error No. Possible cause(s)
ERR 1 No pulse was detected. Ensure the cuff is properly placed on
the inside of you wrist and repeat a measurement
ERR 2 Unnatural pressure impulses influenced the measurement
result. Reason: The arm was moved during the Measurement
(Artefact). Repeat, keeping still and quiet.
ERR 3 Inflation of the cuff takes too long. The cuff is not correctly
seated. Re-position cuff and repeat
ERR 5 The measured readings indicated an unacceptable difference
between systolic and diastolic pressures. Take another read-
ing following directions carefully. Contact your
doctor if you continue to get unusual readings.
HI The cuff pressure is too high. Relax for 5 minutes and
repeat the measurement.*
LO The pulse is too low (less than 40). Repeat the measure-
ment.*
* If this or any other problem occurs repeatedly, please consult
your doctor.
16

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Microlife BP3NC1-1W and is the answer not in the manual?

Microlife BP3NC1-1W Specifications

General IconGeneral
BrandMicrolife
ModelBP3NC1-1W
CategoryBlood Pressure Monitor
LanguageEnglish

Related product manuals