Dear Customer,
This nebuliser is an aerosol therapy system suitable for
domestic use. This device is used for the nebulisation of
liquids and liquid medication (aerosols) and for the treat-
ment of the upper and lower respiratory tract.
If you have any questions, problems or want to order
spare parts please contact your local Microlife-Customer
Service. Your dealer or pharmacy will be able to give you
the address of the Microlife dealer in your country. Alter-
natively, visit the internet at www.microlife.com where
you will find a wealth of invaluable information on our
products.
Stay healthy – Microlife AG!
1. Explanation of symbols
Electronic devices must be disposed of in accor-
dance with the locally applicable regulations, not
with domestic waste.
Read the instructions carefully before using this
device.
Type BF applied part
2. Important Safety Instructions
Follow instructions for use. This document provides
important product operation and safety information
regarding this device. Please read this document
thoroughly before using the device and keep for
future reference.
This device may only be used for the purposes
described in these instructions. The manufacturer
cannot be held liable for damage caused by incorrect
application.
Retain instructions in a safe place for future reference.
Do not operate the unit in presence of any anesthetic
mixture inflammable with oxygen or nitrogen protoxide.
This device is not suitable for anaesthesia and lung
ventilation.
This device should only be used with original acces-
sories as shown in these instructions.
Do not use this device if you think it is damaged or
notice anything unusual.
Never open this device.
This device comprises sensitive components and
must be treated with caution. Observe the storage
and operating conditions described in the «Technical
Specifications» section.
Protect it from:
- water and moisture
- extreme temperatures
- impact and dropping
- contamination and dust
- direct sunlight
- heat and cold
Comply with the safety regulations concerning the
electrical devices and in particular:
- Never touch the device with wet or moist hands.
- Place the device on a stable and horizontal
surface during its operation.
- Do not pull the power cord or the device itself to
unplug it from the power socket.
- The power plug is a separate element from the
grid power; keep the plug accessible when the
device is in use.
Before plugging in the device, make sure that the
electrical rating, shown on the rating plate on the
bottom of the unit, corresponds to the mains rating.
In case the power plug provided with the device does
not fit your wall socket, contact qualified personnel for
a replacement plug with that of a suitable one. In
general, the use of adapters, simple or multiple, and/
or extension cables is not recommended. If their use
is indispensable, it is necessary to use types
complying with safety regulations, paying attention
that they do not exceed the maximum power limits,
indicated on adapters and extension cables.
Do not leave the unit plugged in when not in use;
unplug the device from the wall socket when it is not
in operation.
The installation must be carried out according to the
instructions of the manufacturer. An improper instal-
lation can cause damage to persons, animals or
objects, for which the manufacturer cannot be held
responsible.
Do not replace the power lead of this device. In case of
a power lead damage, contact a technical service center
authorized by the manufacturer for its replacement.
The power supply cord should always be fully
unwound in order to prevent dangerous overheating.
Before performing any maintenance or cleaning
operation, turn off the device and disconnect the plug
from the main supply.
Only use the medication prescribed for you by your
doctor and follow your doctor's instructions with
regard to dosage, duration and frequency of the
therapy.
Depending on the pathology, only use the treatment
that is recommended by your doctor.
Only use the nose piece if expressly indicated by your
doctor, paying special attention NEVER to introduce
the bifurcations into the nose, but only positioning
them as close as possible.
Check in the medicine instruction leaflet for possible
contraindications for use with common aerosol
therapy systems.
Do not position the equipment so that it is difficult to
operate the disconnection device.
For greater hygienic safety, do not use the same
accessories for more than one person.
Never bend the nebuliser over 60°.
Do not use this device close to strong electromagnetic
fields such as mobile telephones or radio installations.
Keep a minimum distance of
3.3
m from such devices
when using this device.
Ensure that children do not use this device unsu-
pervised; some parts are small enough to be swal-
lowed. Be aware of the risk of strangulation in
case this device is supplied with cables or tubes.
Use of this device is not intended as a substitute
for a consultation with your doctor.
3. Preparation and Usage of this Device
Prior to using the device for the first time, we recommend
cleaning it as described in the section «Cleaning and
Disinfecting».
1. Assemble the nebuliser kit AK. Ensure that all parts
are complete.
2. Fill the nebuliser with the inhalation solution as per
your doctor’s instructions. Ensure that you do not
exceed the maximum level.
3. Connect the nebuliser with the air hose 6 to the
compressor 1 and plug the power lead 2 into the
socket (230V 50 Hz AC).
4. To start the treatment, set ON/OFF switch 3 into the
«I» position.
- The mouthpiece gives you a better drug delivery
to the lungs.
- Choose between adult 8 or child face mask 9
and make sure that it encloses the mouth and
nose area completely.
- Use all accessories including the nose piece AL
as prescribed by your doctor.
5. During inhalation, sit upright and relaxed at a table
and not in an armchair, in order to avoid compressing
your respiratory airways and impairing the treatment
effectiveness. Do not lie down while inhaling. Stop
inhalation if you feel unwell.
6. After completing the inhalation period recommended by
your doctor, switch the ON/OFF switch
3
to position
«O»
to turn off the device and unplug it from the socket.
7. Empty the remaining medication from the nebuliser
and clean the device as described in the section
«Cleaning and Disinfecting».
This device was designed for intermittent use
of 30 min. On / 30 min. Off. Switch off the
device after 30 min. use and wait for another
30 min. before you resume treatment.
The device requires no calibration.
No modification to the device is permitted.
4. Cleaning and Disinfecting
Thoroughly clean all components to remove medication
residuals and possible impurities after each treatment.
Use a soft and dry cloth with non-abrasive cleaners to
clean the compressor.
Make sure that the internal parts of the device
are not in contact with liquids and that the power
plug is disconnected.
Cleaning and disinfecting of the accessories
Follow carefully the cleaning and disinfecting instructions
of the accessories as they are very important to the
performance of the device and success of the therapy.
Before and after each treatment
Disassemble the nebuliser
5
by turning the top counter-
clockwise and remove the medicine conduction cone.
Wash the components of the disassembled nebuliser, the
mouthpiece
7
and the nose piece
AL
by using tap water;
dip in boiling water for
5
minutes. Reassemble the nebu-
liser components and connect it to the air hose connector,
switch the device on and let it work for
10-15
minutes.
Wash masks and air tube with warm water.
Only use cold disinfecting liquids following the
manufacturer’s instructions.
Do not boil nor autoclave the air tube and masks.
5. Maintenance, Care, and Service
Order all spare parts from your dealer or pharmacist, or
contact Microlife-Service (see foreword).
Replacement of the nebuliser
Replace the nebuliser 5 after a long period of inactivity,
in cases where it shows deformities, breakage, or when
the vaporiser head 5-a is obstructed by dry medicine,
dust, etc. We recommend to replace the nebuliser after
a period between 6 months and 1 year depending on the
usage.
Only use original nebulisers!
Replacement of the air filter
In normal conditions of use, the air filter AM must be
replaced approximately after 500 working hours or after
each year. We recommend to periodically check the air
filter (10 - 12 treatments) and if the filter shows a grey or
brown colour or is wet, replace it. Extract the filter and
replace it with a new one.
Do not try to clean the filter for reusing it.
The air filter shall not be serviced or maintained
while in use with a patient.
Only use original filters! Do not use the device
without filter!
6. Malfunctions and Actions to take
The device cannot be switched on
Ensure the power lead 2 is correctly plugged into
the socket.
Ensure the ON/OFF switch 3 is in the position «I».
Make sure that the device has been operating within
operating limits indicated in this manual (30 min On /
30 min Off).
The nebuliser functions poorly or not at all
Ensure the air hose 6 is correctly connected at both
ends.
Ensure the air hose is not squashed, bent, dirty or
blocked. If necessary, replace with a new one.
Ensure the nebuliser 5 is fully assembled and the
colored vaporiser head 5-a is placed correctly and
not obstructed.
Ensure the required medication has been added.
7. Guarantee
This device is covered by a 5 year guarantee from the
date of purchase. The guarantee is valid only on presen-
tation of the guarantee card completed by the dealer
(see back) confirming date of purchase or the receipt.
The guarantee covers only the compressor. The
replaceable components like nebuliser, masks,
mouthpiece, air hose, and filters are not included.
Opening or altering the device invalidates the guarantee.
The guarantee does not cover damage caused by
improper handling, accidents or non-compliance with
the operating instructions.
8. Technical Specifications
This device complies with the requirements of the
Medical Device Directive 93/42/EEC.
Class II device as regards protection against electric
shocks.
Nebuliser, mouthpiece and masks are type BF applied
parts.
The technical specifications may change without
prior notice.
Date of production: first 8 digits of the serial number of
the device. First 4 digits: year / 5th and 6th digit: month /
7th and 8th digit: day of production.
Portable Aerosol Compressor Nebuliser NEB 200 Microlife
EN
1
Piston compressor
2
Power lead
3
ON/OFF Switch
4
Air filter compartment
5
Nebuliser
-a: Vaporiser head
6
Air hose
7
Mouthpiece
8
Adult face mask
9
Child face mask
AT
Replacing air filter
AK
Assembling nebuliser kit
AL
Nose piece
AM
Air filter
Class II equipment
Serial number
Reference number
Manufacturer
ON
OFF
Protection against solid foreign objects and
harmful effects due to the ingress of water
CE Marking of Conformity
S
N
N
IP21
0123
Nebulisation rate:
0.4 ml/min. (NaCI 0.9%)
Particle size:
63% < 5 µm
2.6
µm (MMAD)
Max. free air flow: 15 l/min.
Operating air flow: 5.31 l/min.
Acoustic noise level:
52 dBA
Power source:
230V 50 Hz AC
Power lead length:
1.6 m
Nebuliser capacity:
min. 2 ml; max. 8 ml
Residual volume:
0.8 ml
Operating limits:
30 min. On / 30 min. Off
Operating conditions:
10 - 40 °C / 50 - 104 °F
10 - 95 % relative maximum
humidity
700 - 1060 hPa Atmospheric
pressure
Storage and shipping
conditions:
-20 - +60° C / -4 - +140 °F
10 - 95 % relative maximum
humidity
700 - 1060 hPa Atmospheric
pressure
Weight:
1306 g
Dimensions:
103 x 160 x 140 mm
Included: Piston compressor, nebuliser,
air hose, adult face mask, child
face mask, mouthpiece, nose
piece, 5 pcs. of air filter, 2 pcs. of
elastic bands for masks,
carrying bag, instruction
manual, guarantee card
IP Class:
IP21
Reference to
standards:
EN 13544-1; EN 60601-1;
EN 60601-2; EN 60601-1-6;
IEC 60601-1-11
Expected service life: 1000 hours