MCAC-VTSM-2015-10      R410A All DC Inverter V5 X Series 50Hz 
Installation                                                                                                                                                         117 
11. Vacuum Drying 
11.1 Purpose and main points of vacuum drying 
11.1.1 Purpose of vacuum drying 
1.  Remove  moisture  from  the  system  to  prevent  ice-blockage  and  copperizing.  Ice-blockage  will  cause 
abnormal operation, while copperizing may damage compressor. 
2. Eliminating the non-condensable gas of system prevents oxidizing of internal components, system pressure 
fluctuation, and low heat exchange performance. 
3. Detect leak source from reverse rotate. 
11.1.2 Selection of vacuum pump 
1. Vacuuming suction must be less than -756mmHg. 
2. The discharge of vacuum pump is over 4L/s. 
3. The precision of vacuum pump is over 0. 02mmHg. 
R410A system caution: 
After  the  vacuuming process  of R410A  refrigerant circulation  system is  complete, vacuum  pump  will  stop 
running. As a result of suction in the in the air conditioning system, vacuum pump lubricant will seep back into 
the system. This, same situation will also occur if vacuum pump suddenly stops during operation. This will 
cause different oils mix, leading to system malfunction, it is therefore recommended to use one-way valve to 
prevent reverse flow of oil in vacuum pump. 
11.1.3 Vacuum drying for piping 
Vacuum drying: Use of vacuum pump effectively transfers moisture in lines from liquid to gas state. Typical 
boiling point for water is 100
o
C the vacuum pump reduces the internal pressure of the piping system thus 
reducing the boiling point of water to room temperature. This process evaporates all moisture in the piping 
system. 
Boiling Point of 
Water (°C) 
Boiling Point of 
Water (°C) 
11.2 Operation procedure 
There are two vacuum drying methods: common and special. 
11.2.1 Ordinary vacuum drying 
1) Connect the pressure gauge to the infusing mouth of gas and liquid pipe, keep vacuum pump running for at 
least 2 hours, ensures vacuum degree o is below -755mmHg. 
2) If the vacuum degree is not below -755mmHg after 2 hours of drying, system will continue drying for one 
hour. 
3) If the vacuum degree is not be below -755mmHg after 3 hours of drying, check the system for leaks. 
4) Vacuum placement test: when the vacuum degree reaches -755mmHg, keep for 1 hour. If the vacuum 
gauge indicator does not increase, the process is complete. If the vacuum gauge increases, it indicates of 
moisture or a leak. 
5) Vacuum drying shall be conducted from liquid and gas pipe simultaneously. There are a lot of functional 
parts like valves, which could shut down the gas flow midway. 
11.2. 2 Special vacuum drying 
Special vacuum drying method shall be adopted when: