OPERATION
y The meter is supplied complete with a 9V battery.
Slide o the battery compartment cover on the back
of the meter. Install the battery into the battery clip
connector while observing polarity.
y Always remove the electrode protective cap before
taking any measurement. If the electrode has been
left dry, soak the tip (bottom 2.5 cm) in rinse solution
(M10000B) for a few minutes to reactivate it.
y Connect the pH electrode to the BNC socket on the
top of the meter.
y Make sure that the meter has been calibrated
before taking any measurements (see Calibration
Procedure).
y Immerse the tip (2.5 cm) of the pH electrode into the
sample.
y Turn the instrument on by pressing the ON/OFF key.
y Allow the reading to stabilize before taking
measurements.
y After measurements, switch the meter o and store
the electrode with a few drops of storage solution
(SE-220) in the protective cap.
CALIBRATION PROCEDURE
y Remove the protective cap from the electrode.
y Rinse tip o with distilled water or a small quantity
of pH 7.
y Immerse the electrode tip into a new sachet of pH7
calibration solution and allow the reading to stabilize.
y Adjust the pH7 (OFFSET) calibration trimmer to display
“7.0pH”.
y Open a new sachet of pH4 calibration solution and
use a small quantity to rinse the electrode.
y Immerse the electrode tip inside the pH4 sachet and
allow the reading to stabilize.
y Adjust the pH4/pH10 (SLOPE) calibration trimmer to
read “4.0 pH”.
y The calibration is now complete and the meter is
ready to take measurements.
y It is suggested the meter be calibrated at least once a
month, after a prolonged period of inactivity, and after
electrode replacement.
BATTERY REPLACEMENT
When the battery becomes weak the meter will
display “ “.
When the low battery indicator appears, the battery has
only about 50 hours of working time left. A low battery will
result in unreliable measurements.
Prompt battery replacement is required. Battery
replacement must only take place in a non-hazardous
area using an alkaline 9V battery.
Turn the meter o, slide the battery compartment
cover located at the rear of the meter o and replace
the 9V battery with a new one. Make sure the battery
contacts are fully engaged in the connector, seat the
battery in its compartment and replace the cover.
OPTIONAL ACCESSORIES
M10004B
pH 4.01 buer solution,
20 mL sachet (25 pcs)
M10007B
pH 7.01 buer solution,
20 mL sachet (25 pcs)
M10010B
pH 10.01 buer solution,
20 mL sachet (25 pcs)
M10000B
Electrode rinse solution,
20 mL sachet (25 pcs)
MA9015
Electrode storage solution,
220 mL bottle