A.6 Interface Specifications
A.7 Battery Specifications
USB connector 1, USB 2.0
micro USB connector 1, supports Windows 7 or above operating system
Network connector 1, connects the Wi-Fi or cellular (2G/3G/4G) network.
Multifunction connector 1, connects the CPR sensor.
Battery type Disposable battery
Battery voltage 12V
Battery capacity 4200mAh
Equipment configured without
the screen
Operating
time
Testing condition
15 hours The equipment is powered by a new battery at 20 °C± 5
°C of ambient temperature, wireless function off, not
performing defibrillation charges or discharges, voice
volume set to low.
300 200J
discharges
The equipment is powered by a new battery at 20 °C± 5
°C of ambient temperature, wireless function off, voice
volume set to low, with one minute of CPR between
discharges
190 360J
discharges
510 150 J
discharges
The equipment is powered by a new battery at 20 °C± 5
°C of ambient temperature, wireless function off, voice
volume set to low, with three discharges every minute
400 200J
discharges
200 360J
discharges
Equipment configured with the
screen
12 hours The equipment is powered by a new battery at 20 °C± 5
°C of ambient temperature, wireless function off, not
performing defibrillation charges or discharges, voice
volume set to low.
270 200J
discharges
The equipment is powered by a new battery at 20 °C± 5
°C of ambient temperature, wireless function off, voice
volume set to low, with one minute of CPR between
discharges
170 360J
discharges
450 150J
discharges
The equipment is powered by a new battery at 20 °C± 5
°C of ambient temperature, wireless function off, voice
volume set to low, with three discharges every minute
350 200J
discharges
200 360J
discharges
Battery fuel gauge (for
equipment configured with the
screen)
Battery symbol on the display indicating the current battery level

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneHeart C Series and is the answer not in the manual?

Mindray BeneHeart C Series Specifications

General IconGeneral
BrandMindray
ModelBeneHeart C Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals