EasyManuals Logo

Mindray Passport 8 Operator's Manual

Mindray Passport 8
284 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #120 background imageLoading...
Page #120 background image
11-8 Passport 12/Passport 8 Operator’s Manual
Symptoms Possible Cause Correction Action
No SpO
2
waveform Waveform not selected to display Switch the SpO
2
monitoring function on as
described in 3.12.1 Switching the Parameters
On/Off.
Cable or sensor not plugged in Check that the cable is properly connected and
sensor securely applied.
Low amplitude SpO
2
signal SpO
2
sensor on same limb as cuff Check that the sensor is properly applied. Change
the application site if necessary.
Patient has poor perfusion Change the application site.

Table of Contents

Other manuals for Mindray Passport 8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Passport 8 and is the answer not in the manual?

Mindray Passport 8 Specifications

General IconGeneral
BrandMindray
ModelPassport 8
CategoryMedical Equipment
LanguageEnglish

Related product manuals