A-12 Passport 12/Passport 8 Operator’s Manual
PR from NIBP Module
Measurement range 40 to 240 bpm
Resolution 1 bpm
Accuracy ±3bpm or ±3%, whichever is greater
PR from IBP Module
Measurement range 25 to 350 bpm
Resolution 1 bpm
Accuracy ±1 bpm or ±1%, whichever is greater
Refreshing rate 1 s
A.7.5 NIBP
Standards
Meet standards of EN60601-2-30/IEC60601-2-30, EN1060-1, EN1060-3, SP10 and
EN1060-4
Technique Oscillometry
Mode of operation Manual, Auto and STAT
Auto mode repetition intervals
1 min, 2 min, 2.5 min, 3 min, 5 min, 10 min, 15 min, 20 min, 30 min, 1 h, 1.5 h, 2 h, 3 h,
4 h, 8 h
STAT mode cycle time 5 min
Max measurement time
Adult, pediatric: 180 s
Neonate: 90 s
Measurement ranges
(mmHg)
Adult Pediatric Neonate
Systolic: 40 to 270 40 to 200 40 to 135
Diastolic: 10 to 210 10 to 150 10 to 100
Mean: 20 to 230 20 to 165 20 to 110
Accuracy
Max mean error: ±5 mmHg
Max standard deviation: 8 mmHg
Resolution 1mmHg
Initial cuff inflation pressure range
(mmHg)
Adult:
Pediatric:
Neonate:
80 to 280
80 to 210
60 to 140
Default initial cuff inflation pressure
(mmHg)
Adult:
Pediatric:
Neonate:
160
140
90
Software overpressure protection
Adult:
Pediatric:
Neonate:
297±3 mmHg
240±3 mmHg
147±3 mmHg
Hardware overpressure protection
Adult:
Pediatric:
Neonate:
≦330 mmHg
≦330 mmHg
≦165 mmHg