Passport 12/Passport 8 Operators Manual A-17
A.7.10 AG (For Passport 12 only)
Standards Meet standard of ISO 21647
Technique Infrared absorption
Warm-up time
Iso accuracy mode:
Full accuracy mode:
≤ 45 s
≤ 10 min
Sample flow rate
Adult, pediatric:
Neonate:
Accuracy:
120, 150, 200 ml/min
70, 90, 120 ml/min
±10 ml/min or ±10%, whichever is greater
Measurement range
CO
2
:
O
2
:
N
2
O:
Des:
Sev:
Enf:
Iso:
Hal:
awRR:
0 to 30%
0 to 100%
0 to 100%
0 to 30%
0 to 30%
0 to 30%
0 to 30%
0 to 30%
2 to 100 rpm
Resolution
CO
2
:
awRR:
1 mmHg
1 rpm
Full accuracy
Gases Range (%
REL
) Accuracy (%
ABS
)
CO
2
0 to 1
1 to 5
5 to 7
7 to 10
>10
±0.1
±0.2
±0.3
±0.5
Not specified
N
2
O
0 to 20
20 to 100
±2
±3
O
2
0 to 25
25 to 80
80 to 100
±1
±2
±3
Des
0 to 1
1 to 5
5 to 10
10 to 15
15 to 18
>18
±0.15
±0.2
±0.4
±0.6
±1
Not specified
Sev
0 to 1
1 to 5
5 to 8
>8
±0.15
±0.2
±0.4
Not specified
Enf, Iso, Hal
0 to 1
1 to 5
>5
±0.15
±0.2
Not specified
awRR
2 to 60 rpm
>60 rpm
±1 rpm
Not specified