Passport 12/Passport 8 Operator’s Manual C-5
C.1.2 RESP
Item Name
Configurable
Default
In Config Mode In Monitor Mode
Alarm Yes Yes On
Alm Lev Yes Yes Med
RR High Yes Yes
Adu, Ped:
Neo:
30
100
RR Low Yes Yes
Adu, Ped:
Neo:
8
30
Apnea Delay Yes Yes
Adu, Ped:
Neo:
20
15
Lead Yes Yes
Adu, Ped: Auto
Neo: II
Gain Yes Yes X2
Sweep Yes Yes 6.25 mm/s
Detection Mode Yes Yes Auto
RR Source No Yes Auto
C.1.3 PR
Item Name
Configurable
Default
In Config Mode In Monitor Mode
Alarm Yes Yes On
Alm Lev Yes Yes Med
HR/PR High Yes Yes
Adu: 120
Ped: 160
Neo: 200
HR/PR Low Yes Yes
Adu: 50
Ped: 75
Neo: 100
PR Source Yes Yes SpO
2
Alm Source Yes Yes HR
Beat Vol Yes Yes
General, OR: 2
ICU, NICU, CCU: 1