C-10 Passport 12/Passport 8 Operator’s Manual
C.1.9 CO
2
Item Name
Configurable
Default
In Config Mode In Monitor Mode
Alarm Yes Yes On
Alm Lev Yes Yes Med
Apnea Delay Yes Yes
Adu, Ped:
Neo:
20
15
Operating Mode Yes Yes Measure
Sweep Yes Yes 6.25 mm/s
Scale (mmHg) Yes Yes 50
RR Source No Yes Auto
Sidestream CO
2
Setup
Flow Rate Yes Yes
Adu,
Ped:
Neo:
120 ml/min
100 ml/min
70 ml/min
BTPS Compen Yes Yes Off
N
2
O Compen Yes Yes 0
O
2
Compen Yes Yes
General: 21
OR: 100
ICU, NICU, CCU: 21
Des Compen Yes Yes 0
Microstream CO
2
Setup
BTPS Compen Yes Yes Off
Max Hold Yes Yes 20 s
Auto Standby (min) Yes Yes 0
Alarm Limits
EtCO
2
High (mmHg) Yes Yes
Adu, Ped:
Neo:
50
45
EtCO
2
Low (mmHg) Yes Yes
Adu, Ped:
Neo:
25
30
FiCO
2
High (mmHg) Yes Yes Adu, Ped, Neo: 4
RR High Yes Yes
Adu, Ped:
Neo:
30
100
RR Low Yes Yes
Adu, Ped:
Neo:
8
30

Table of Contents

Other manuals for Mindray Passport 8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Passport 8 and is the answer not in the manual?

Mindray Passport 8 Specifications

General IconGeneral
BrandMindray
ModelPassport 8
CategoryMedical Equipment
LanguageEnglish

Related product manuals