15. Specification | 15.3 - Effect of ambient conditions | 140
15.3 Effect of ambient conditions
15.3.1 Ambient conditions
Temperature range
Ambient temperature for oper
ation
-10…+55°C
Power-on temperature >0°C
Limits for storage/transport -20…+70°C
Humidity <90%, non-condensing (acc. to
IEC60068-2)
Protection class IP20
Height <2000m
Vibrations The device should not be exposed to
strong vibrations.
15.3.2 Electromagnetic Compatibility (EMC)
All data in compliance with EN 61326 industrial section
High frequency electromagnetic
fields (80…1000MHz)
EN61000-4-3 10V/m
High frequency electromagnetic
fields (1.4…2.0GHz)
EN61000-4-3 3V/m
High frequency electromagnetic
fields (2.0…2.7GHz)
EN61000-4-3 1V/m
Housing
Electrostatic discharge (ESD) EN61000-4-2 4/8kV
Fast transients (burst) EN61000-4-4 1kV
Peak voltages (surge) 1.2µs EN61000-4-5 1kV
Signal and control
lines
Conducted disturbances by
high frequency coupling (0.15…
80MHz)
EN61000-4-6 3V