EasyManua.ls Logo

Minebea Intec MiNexx-link-E PR 5215/10 - RF Interference Suppression; Weighing Electronics Data; Load Cells

Default Icon
150 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
15. Specification | 15.4 - Weighing electronics data | 141
Fast transients (burst) EN61000-4-4 2kV
Peak voltages (surge) 1.2/50µs EN61000-4-5 1/2kV
Mains inputs
Conducted disturbances by
high frequency coupling (0.15…
80MHz)
EN61000-4-6 3V
15.3.3 RF interference suppression
Electromagnetic
emission
pursuant to EN61326, Limit classA, for
industrial areas
15.4 Weighing electronics data
15.4.1 Load cells
Load cell type Strain gauge
load cells
6 or 4-wire connection possible.
Supply voltage U
DC
=5V
for I
max
= 66.6mA
for max. 8load cells, each
with 650Ω
4load cells, each with 350Ω
Max. load ≥75Ω

Table of Contents

Related product manuals