Ex.2 In case of net value cancelled 300 g of tare weight set in ex.1
The measurement from -300 to 300 g becomes the first range (scale interval 0.1 g).
The measurement from 300 to 600 g becomes the second range (scale interval 0.5 g).
The measurement from 600 to 700 g becomes the second range (scale interval 1 g).
・ When it is used only within the first range (C function CF-40: OFF), it is not necessary to set the
range.
・ Please make the relations of the scale interval of the each range as the first range<the second
range<the third range.
・ Please make the relations of the each boundary value of each range as the second range≦the third
range≦the weighing capacity.
・ The range upper bound within the range of second is a weighing capacity for the switch by two
ranges.
The range upper bound within the range of third is a weighing capacity for the switch by three
ranges.
・ The comparator setting and the preset tare weight cancellation setting become the scale interval set
in the first range.
6-15. Setting of the stability detection time width in calibration
The stability detection time width in calibration can be set by C function CF-97.
Setting range: 00 ~ 99 Units: 0.1 s
Default is set to [05].
6-16. Digital linearize clear
The compensation data set in the digital linearization can be cleared by the C function CF-98.
If the key is pressed with display, becomes a blinking
display.
Press the key to discontinue the digital linearization at this point.
It becomes a stand-by condition, and the digital linearization is not clear.
is displayed If the key is pressed when displays blinking, and
the operation of digital linearization is completed.