• H*(10) dose equivalent rate according to latest standards
• External alpha, beta, gamma and neutron probes
• Congurable levels for warning and alarm with audible, visual and vibra-
tion indication
• Congurable units: Sv(/h), rem(/h), with external detectors cps, cpm,
dpm, Bq and Bq/cm
2
• Power supply: 2 x AA (R6) size batteries (alkaline or NiMH)
• Operation time with fresh alkaline batteries 600 h in background radia-
tion, radios disabled, display backlight o, LED o
• Real time clock function with 3 hrs backup
• Enclosure class IP67
• iTx versions with internal radio for wireless monitoring
RDS-32/RDS-32iTx
• Detector: energy compensated GM tube
• IEC energy range: 48 keV to 1.8 MeV
• Dose rate measurement range: 0.05 μSv/h to 100 mSv/h
(5 μrem/h to 10 rem/h)
• IEC dose rate range: 0.3 μSv/h to 100 mSv/h
(0.03 mrem/h to 10 rem/h)
• Dose measurement range: 0.1 μSv to 10 Sv (0.01 mrem to 1000 rem)
RDS-32WR/RDS-32iTxWR
• Detector: energy compensated GM tube, Si diode
• IEC Energy range: 55 keV (GM tube) / 65keV* (Si Diode) to 1.8 MeV
• Dose rate measurement range: 0.05 μSv/h to 10 Sv/h
(5 μrem/h to 1000 rem/h)
• IEC dose rate range: 0.3 μSv/h to 10 Sv/h
(0.03 mrem/h to 1000 rem/h)
• Dose measurement range: 0.1 μSv to 10 Sv (0.01 mrem to 1000 rem)
* GM tube to Si diode at 30 mSv/h in increasing eld,
Si diode to GM tube at 10 mSv/h in decreasing eld
RDS-32 Quick Guide, Doc. no. 20967193 v.1.0
Mirion Technologies (RADOS) Oy
COMMON SPECIFICATION