OPERATING MANUAL
MBD-2 (IM-278/U)
Personal Dosimeter
Document # 15-00167
Revision 1
December 2018

Table of Contents

Questions and Answers

  • B
    Bradley CarrilloSep 18, 2025
    What can cause saturation in Mirion Technologies MBD-2 Measuring Instruments detectors?
    • N
      nicholas64Sep 18, 2025
      Saturation in one or more detectors of your Mirion Technologies Measuring Instruments can be caused by: * Overflow Total Dose: the total dose reaching/exceeding 3,750 cGy * Overflow Gamma: the gamma dose reaching/exceeding 2,500 cGy * Overflow Neutron: the neutron dose reaching/exceeding 1,250 cGy
  • P
    Peggy MorganAug 25, 2025
    What does it mean if my Mirion Technologies Measuring Instruments display 'dbAt'?
    • Z
      ztaylorAug 25, 2025
      If 'dbAt' is cycling on the display of your Mirion Technologies Measuring Instruments, it means less than 25% battery life remains. If 'dbAt' is permanently displayed, the battery is exhausted. In both cases, replace the batteries with 3 new CR1025 batteries.
  • J
    Jose ReevesAug 25, 2025
    What to do if the Mirion Technologies MBD-2 device restarts after a battery change?
    • P
      pcooperAug 26, 2025
      If your Mirion Technologies Measuring Instruments device restarts after a battery change, it indicates that internal data memory was found to be corrupt, and the device performed a self-hard reset to clean the memory. If this error message occurs under normal use (without high dose or neutron exposure), remove the device from service.
  • K
    kevin89Aug 27, 2025
    What to do if Mirion Technologies MBD-2 Measuring Instruments show parameter error?
    • J
      John ValenzuelaAug 27, 2025
      If your Mirion Technologies Measuring Instruments device shows a Parameter Error, remove the device from service and return it for Warranty Repair.
  • R
    ricky45Sep 22, 2025
    What to do if Mirion Technologies Measuring Instruments detectors can't report a measurement value?
    • L
      Laura Smith MDSep 22, 2025
      If one or more detectors on your Mirion Technologies Measuring Instruments device cannot report a measurement value, remove the device from service. Remove and replace the battery and observe the display. If the problem persists, return the device for Warranty Repair.
  • M
    Marc TuckerSep 23, 2025
    What to do if Mirion Technologies MBD-2 Measuring Instruments report recombination?
    • G
      garrisonkatherineSep 23, 2025
      If recombination has occurred in an ionization chamber element due to a gamma flash or neutron burst event on your Mirion Technologies Measuring Instruments device, it requires a reset of the device. Remove the device from service.
  • Y
    Yolanda GomezSep 9, 2025
    What to do if Mirion Technologies Measuring Instruments cannot determine reading result position?
    • H
      Heidi PowellSep 10, 2025
      If your Mirion Technologies Measuring Instruments device cannot determine the reading result position in flash memory due to a failure in flash memory access or corruption of memory contents, restart the device by removing and re-inserting the battery. If the error message repeats, remove the device from service.
  • S
    Shannon ReynoldsSep 12, 2025
    What to do if Mirion Technologies MBD-2 Measuring Instruments reading fails due to corruption of result memory?
    • B
      buckleyvictoriaSep 12, 2025
      If the measurement result reading fails on your Mirion Technologies Measuring Instruments device due to corruption of the result memory, restart the device by removing and re-inserting the battery. If the error message repeats, remove the device from service.
  • T
    Tony RojasSep 3, 2025
    What to do if Mirion Technologies MBD-2 Measuring Instruments show low battery voltage?
    • N
      nancymarshallSep 3, 2025
      If your Mirion Technologies Measuring Instruments device shows a low main battery voltage (less than 2200 mV) with no battery icon and a permanent 'PbAt' message, replace the main battery with model CR2450.
  • V
    Vanessa MooreSep 1, 2025
    What to do if Mirion Technologies MBD-2 report exceeding maximum dose?
    • M
      michaelpettySep 1, 2025
      If your Mirion Technologies Measuring Instruments device reports that one or more detector elements has exceeded the maximum dose, it requires a reset of the device. Remove the device from service.

Mirion Technologies MBD-2 Specifications

General IconGeneral
BrandMirion Technologies
ModelMBD-2
CategoryMeasuring Instruments
LanguageEnglish