In this section, “typical” refers to a specification that is within ±20% of a stated value at room
temperature (25°C/77°F).
VAPR ELECTRODE SPECIFICATIONS
Overall Length: 5.7 – 8.9 in (14.5 – 22.5 cm)
Working Length: 3.7 – 6.5 in (9.5 – 16.5 cm)
Shaft Diameter: 0.06 – 0.18 in (1.5 – 4.5 mm)
Shaft Bend Angle: 0 – 30°
Active Tip Orientation: 0 – 90°
Supplied Sterile and For Single-Use Only
Sterilization Method: Irradiation
VAPR HANDPIECE AND CABLE SPECIFICATIONS
Overall Length: 10.5 ft (3.2 m)
Sterilization Method: Steam (134 – 136°C)
Supplied Non-Sterile
20 x Reusable
VAPR GENERATOR SPECIFICATIONS
Dimensions (H x W x D): 3.5 in x 16 in x 14.5 in (9 cm x 41 cm x 36.8 cm)
Weight: 12 lb. (5.6kg) approximately
Transport and Storage Conditions
Ambient Temperature Range: 32 to 122°F (0 to 50°C)
Relative Humidity: 10% to 90%, non condensing
Atmospheric Pressure: 500 to 1060 millibars
Operating Conditions
Ambient Temperature Range: 60 to 104°F (10 to 40°C)
Relative Humidity: 10% to 90%, non condensing
Atmospheric Pressure: 500 to 1060 millibars
Power Supply
Regulation Voltage: 90 – 132V RMS, 198 – 264V RMS
Operating Range: nominal 100 – 120/220 – 240V RMS 50/60Hz
Inlet Fuses: Time lag 5A (T5AH 250V) (VAPR II ONLY)
Inlet Fuses: Time lag 6.3A (T6A3H 250V) (VAPR 3 ONLY)
Leakage Currents: Within limits of Class BF equipment as per IEC 60601-1
Alarm Volume: Adjustable between 40 dB (minimum) and 65dB
(maximum) at 1m. This is an activation signal only.
Classification: Electrical: Class 1 ordinary equipment as per IEC60601-1.
EMC: Group 1 Class A as per IEC60601-1-2.