Mitutoyo ARK-600 - Tester Calibration; Preliminary Test Force Measurement; Test Force Calibration Procedure; Indicator Calibration Steps

Mitutoyo ARK-600
28 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
.
TESTER
CALIBRATION
(Unnecessary
in
ordinary
maintenance
of
tester.
Special
divices
are
required
in
this
calibration.)
To
check
the
load
and
indicator,
set
the
Measurement/Calibration
switch
of
the
tester
side
panel
to
the
"TEST"
position.
8.1
Preliminary
Test
Force
Measurement
Put
a
load
measuring
device
on
the
specimen
base.
Apply
the
preliminary
test
force
and
read
a
load
value
when
all
pilot
lamps
light
up.
(Displayed
value:
“365.0")
8.2
Test
Force
Calibration
1)
P
l
a
ce
a
load
measuring
device
on
the
anvil
(or
table).
Apply
the
preliminary
load,
and
set
the
Measurement
/
Calibation
switch
on
the
side
panel
to
the
"RESET"
position
in
the
same
condition
as
above
(displayed
value:
"365.0"
)
in
order
to
reset
the
displayed
value
to
the
standard
value
(100.0).
2)
Elevate
the
anvil
to
apply
preload
suitable
for
the
load
measuring
device
used.
3)
Press
the
START
switch
to
apply
the
test
force.
Read
displayed
hardness
values
as
related
to
the
load.
For
evaluating
the
load
corresponding
to
another
hardness
value
on
the
display,
modify
the
preload
condition
mentioned
in
the
previous
item,
or
change
the
position
of
the
indenter
by
raising
or
lowering
the
specimen
base
with
the
load
applied,
and
mesure
the
load.
4)
To
release
the
load,
press
the
START
switch
again.
8.3
Indicator
Calibration
1)
Put
a
meter
testing
device
on
the
specimen
base.
The
device
shall
be
regarded
as
the
datum
position.
2)
Elevate
the
anvil
to
apply
the
preliminary
test
force.
3)
Lower
the
spindle
of
the
meter
testing
device,
and
elevate
it
again
to
set
it
to
the
datum
position.
4)
Press
the
RESET
switch
for
approx.
1
second
to
reset
the
displayed
value
to
the
standard
value.
5)
Lower
the
spindle
of
the
meter
testing
device
below
"-30".
Then,
raise
it
a
little
to
measure
the
reading
error
in
the
upward
direction
of
the
indenter.