A. BOILING
Place the four disassembled parts in a pot when boiling point has been reached and allow to boil for 5 minutes. Remove
from water and allow parts to cool and air dry thoroughly before reassembly.
NOTE
:Placeasmallrack,onthebottomofthepottopreventplasticpartsfromcomingincontactwiththehotbottom.
B. MICROWAVE STEAM BAG
The four disassembled parts may be disinfected in the microwave
in a steam cleaning bag (i.e. Quick-Clean
MicroSteam
bag)
following the manufacturer’s instructions. Remove from bag and
allow parts to cool and air dry thoroughly before reassembly.
PositionpartsinbagasshowninFigure4.
C. ISOPROPYL ALCOHOL (70%)
Thefourdisassembledpartsmaybesoakedin70%isopropyl
alcohol for 5 minutes. Rinse well and allow to air dry thoroughly
before reassembly.
D. HYDROGEN PEROXIDE (3%)
Thefourdisassembledpartsmaybesoakedin3%hydrogenperoxidefor30minutes.Rinsewellandallowtoairdry
thoroughly before reassembly.
E. BLEACH (Not recommended for high level disinfection)
Thefourdisassembledpartsmaybesoakedin1:50bleachtowatersolution(approximately1tablespoonbleachin
3¼cupswater)for3minutes.Rinsewellandallowtoairdrythoroughlybeforereassembling.Youmaysubstitutea
respiratoryequipmentdisinfectant(suchasMilton
,Dodie
,ControlIII
orCidex
) for the bleach solution by following
the manufacturer’s instructions for use.
CAUTION: Failure to follow the cleaning or disinfection instructions may have an adverse effect on
device performance.
DISPOSAL
The Aerobika
®
Oscillating PEP device can be disposed of with domestic waste unless this is prohibited by the disposal
regulations prevailing in your state.
NOTES
• SINGLEPATIENTUSE.
• Productshouldbereplacedafter6monthsofuseorimmediatelyifdamaged.
• Thisproducthasnotbeenmade(ormanufactured)withBPA(BisphenolA)ornaturalrubberlatex.
• Donotsharethismedicaldevice.
• Ifyouareunabletoexhalethroughthedevice,checktheresistancesettingortrycleaningthedevicethoroughly.
• StoretheAerobika
®
OscillatingPEPdeviceinadry,dust-freeplaceavoidingcontinuousexposuretosunlight.
• Storageandoperatingrange15°C-45°C(59°F-113°F)atambientto85%relativehumidity.
FIGURE 4
FIGURA 1
Lengüetasde
sujeción
Orificio para el
nebulizador
Boquilla
Una versión electrónica de estas instrucciones de uso puede descargarse en www.monaghanmed.com
Parte superior
Cartucho de la
válvula
Parte inferior
Instrucciones de uso
Indicador de resistencia
UNCONTROLLED WHEN PRINTED

Other manuals for Monaghan Aerobika

Related product manuals