EasyManua.ls Logo

Monnit ALTA - Page 13

Monnit ALTA
19 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Creating a Calibration Certicate
Creating a calibration certicate will mask the calibration tab from those who should not
have permissions to adjust these settings. Permissions for self-certifying a calibration must
be enabled in user permissions.
Directly below the calibrate button is the selection to "Create Calibration Certicate.
A. The Calibration Facility Field will be lled. Select the dropdown menu to change your
facility.
B. The date for "Certicate Valid Until" must be set one day in the future after the date
contained in the "Date Certied" eld.
C. "Calibration Number" and "Calibration Type" are unique values to your certicate.
D. Choose the "Save" button before moving on.
When the new certicate is accepted, the Calibration tab will change to a Certicate tab.
You will still be able to edit the certicate by choosing the Certicate Tab and navigating
down to "Edit Calibration Certicate."
The tab will revert back to "Calibrate" after the period for the certicate ends.
PAGE 11