Instructions for Use 2018-04-11
3 Before and After Use
● Use the Veraview X800 under the following conditions:
Temperature: +10°C to +35°C (+50°F to +95°F)
Humidity: 30% to 75% (without condensation);
Atmospheric Pressure: 70 kPa to 106 kPa
* If an accident occurs, the equipment must not be used until repairs have been completed by a qualified
and trained technician authorized by the manufacturer.
* Have patients remove glasses, earrings and any other accessories that might interfere with the exposure.
* Before use make an initial inspection to make sure the unit will operate properly and safely.
3.1 Operation Conditions
3.2 Set Up
3.2.1 Start i-Dixel WEB
Start the i-Dixel or i-Dixel WEB software (hereinafter referred to
as “i-Dixel WEB”).
1. Start i-Dixel or i-Dixel WEB
Enter the patient ID in the search field and click the search icon (A).
To use the patient list, select a patient and then double-click it.
2. Retrieve a registered patient
To begin, register a patient. Click the icon B to register a
new patient. Fill out the form that appears and then click the
Register button.
● Register a New Patient
An image list will appear.
* Patient ID, Name and
Sex are required items.
i-Dixel WEB
i-Dixel
A
B
i-Dixel WEB
A
B
i-Dixel