Ambient Condition for Calibration
- Ambient temperature: 23°C ± 1°C;
- Relative humidity: 45 to 75% RH;
Preheating: standard instrument should be
preheated to specified time;
Put this Instrument in calibration
environment for 24 hours, then connect to the
power, and set it in non auto- power-off state, and
preheat for half an hour.
Note
Power supply for calibrator: during the
calibration, it is recommendable to use 2
1.5VLR6batteries.
Operating Output Calibration
Operating calibration in order of items and
calibration points listed in Table 10-2:
Table 10-2
Item
No.
Output Range
Calibration Point
1
DCA/20mA
0
0 FS
FS
1) In power-off state, press the keysOUTPUT
/ INPUTandpowersimultaneously to
power on, and release theOUTPUT /
INPUTkey only after the LCD displays all
the contents. The Instrument enters into the
calibration mode, and the symbols CAL 0
appear in the LCD;