NAG/IM/MTC650/01/22 25
Choose the read-out scale (°C/°F) you require by
activating the key
Set the required temperature using the numeric
keys and press ENTER.
The heater block will heat / cool to SET
temperature.
Wait until the required SET temperature is equal
to the READ temperature.
When the SET and the READ temperature have
been equal for about 10-20 minutes, a beep
audio signal will be heard and a tick mark will
appear in the display. This indicates that you
have achieved the stable thermo well
temperature. Now you can take the readings.
Connect the RTD probes to the corresponding
terminals (loop supply) in the front plate of the
instrument.
Display shows the bath temperature as the
INPUT parameter which corresponds to the READ
temperature.
Note: If you select mvL or mvH while using T/C, add the
voltage corresponding to the ambient temperature with the
INPUT voltage which is shown in the display.