2. SPECIFICATION
Range 50°C to 650°C
Resolution 0.1°C
Accuracy ±0.6°C
Stability ±0.05°C
Stabilisation Time 15 to 20 Minutes
Temperature Readout °C / °F Selectable
Well Diameter 25 mm
Immersion Depth 110 mm
Heating Time (Ambient to
max.)
20 Minutes approx.
Cooling Time max. (to
100°C)
42 Minutes
Switch Test facility Provided
Interchangeable
Thermowells
Single hole to suit 1/4" & 1/2" Probes.
Other Sizes / Multihole available as
optional.
Housing / Mounting Bench-top
Power Supply 230V AC, 50 Hz
Instrument Dimensions (L x
D x H)
160 x 365 x 350 mm
Instrument Weight
(approx.)
11 Kg.
Measurement Capability Thermocouples ISA Types J, K, R, S, N, B, T
RTD Pt100 IEC Std.
mA 0 to 25 mA
mV 0 to 100 / 1000 mV
Volt 0 to 10 V
Resistance 0 to 500Ω / 3.5 KΩ