NI 9201/9221 14 ni.com
Input range
NI 9201....................................... ±10 V
NI 9221....................................... ±60 V
Operating voltage ranges
1
Overvoltage protection
(channel-to-COM) ............................ ±100 V
1
Refer to the Safety Guidelines section for more information about safe operating
voltages.
Module
Measurement Voltage,
Channel-to-COM
Maximum Voltage,
Channel-to-Earth Ground or
COM-to-Earth Ground
Min
(V)
Typ
(V)
Max
(V)
Screw
Terminal
DSUB
NI 9201 ±10.3 ±10.53 ±10.8 250 V
rms
±60 VDC
NI 9221 ±61.4 ±62.50 ±63.8
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com