Questions and Answers

  • C
    courtneyrobbinsAug 11, 2025
    How to fix National Instruments PXIe-4468 PXIe-446x not appearing in MAX?
    • N
      Nicole AndradeAug 11, 2025
      If the National Instruments Control Unit PXIe-446x module is not showing up in Measurement & Automation Explorer (MAX), try the following: * In MAX, expand 'Devices and Interfaces'. * Expand the 'Chassis' tree and refresh the list by pressing F5. * If it's still not listed, power off the system, ensure the hardware is correctly installed, and restart. * Check the Device Manager to verify the PXIe-446x appears there under an NI entry. If there are error conditions, reinstall the NI-DAQmx driver.
  • R
    robingilbertAug 22, 2025
    What to do if the ACCESS LED is off when the National Instruments Control Unit chassis is on?
    • D
      Daniel PowellAug 22, 2025
      If the ACCESS LED is off when the National Instruments Control Unit chassis is on, it could indicate a problem with the chassis power rails, the hardware module, or the LED itself. First, verify that the PXIe-446x appears in MAX. Then: 1. Disconnect any signals from the module front panels. 2. Power off the chassis. 3. Remove the module and inspect it for damage. Do not reinstall a damaged module. 4. Install the module in a different chassis slot. 5. Power on the chassis again. 6. Check if the module appears in MAX. Reset the module in MAX and perform a self-test.
  • M
    Molly DominguezAug 31, 2025
    What to do if National Instruments Control Unit PXIe-446x fails the self-test?
    • B
      bakerthomasAug 31, 2025
      If the National Instruments Control Unit PXIe-446x fails the self-test, try the following steps: 1. Restart the system. 2. Launch MAX and perform the self-test again. 3. Power off the chassis. 4. Reinstall the failed module in a different slot. 5. Power on the chassis. 6. Perform the self-test again.

National Instruments PXIe-4468 Specifications

General IconGeneral
ADC Resolution24 bits
Maximum Sample Rate204.8 kS/s
Input CouplingAC, DC
Input Voltage Range±10 V
Output Voltage Range±10 V
PXI PlatformPXIe
ConnectivityPXI Express
Dynamic Range110 dB
THD+N-100 dB
InterfacePXI Express
Connector TypeBNC
Operating Temperature Range0 °C to 55 °C
DAC Resolution24-bit
Signal ConditioningProgrammable gain
Storage Temperature Range-20 °C to 70 °C
Relative Humidity Range10% to 90% noncondensing