EasyManua.ls Logo

National Instruments PXIe-4468 - 2. 25 B26 B Set the DMM to the AMPS function and the 10 mA range for the 4 mA and 10 mA test points, and the 100 mA range for the 20 mA test point.; 1. 24 B25 BConnect the AI channel under test to the DMM as shown in Figure 3 or Figure 4.; 3. 26 BCreate and configure an AI Voltage task using the values in Table 10.

National Instruments PXIe-4468
40 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PXIe-4468 Calibration Procedure
18 ni.com/services
Figure 4. Initial Connection for IEPE Current-mXLR
1.
DMM
3.
BNC
mXLR Cable
2.
Banana
BNC Adapter
4.
PXIe
-4468 with mXLR
Verification Procedure
Complete the following procedure to verify the IEPE current.
Repeat 2 times, once for each channel.
1. Connect the AI channel under test to the DMM as shown in Figure 3 or
Figure 4.
Repeat 3 times, once for each IEPE current setting.
2. Set the DMM to the AMPS function and the 10 mA range for the 4 mA
and 10 mA test points, and the 100 mA range for the 20 mA test point.
3. Create and configure an AI Voltage task using the values in Table 10.
Table 10: IEPE Current Verification Configuration
Configuration Value
Physical Channels Dev_name/aix
Gain* -20 (dB)
Coupling AC

Table of Contents

Other manuals for National Instruments PXIe-4468

Related product manuals