PXIe-4468 Calibration Procedure
10 ni.com/services
Figure 1. Initial Connection for AI Offset-BNC
1.
Calibrator
3.
BNC Cable
2.
Banana
BNC Adapter
4.
-4468 with BNC
Figure 2. Initial Connection for AI Offset-mXLR
1.
Calibrator
3.
mXLR Cable
2.
Banana
BNC Adapter
4.
-4468 with mXLR
Verification Procedure
Complete the following procedure to verify the AI offset accuracy.
Repeat 2 times, once for each channel.
1. Connect the AI channel under test to the Calibrator as shown in
Figure 1 or Figure 2.