C
courtneyrobbinsAug 11, 2025
How to fix National Instruments PXIe-4468 PXIe-446x not appearing in MAX?
- NNicole AndradeAug 11, 2025
If the National Instruments Control Unit PXIe-446x module is not showing up in Measurement & Automation Explorer (MAX), try the following: * In MAX, expand 'Devices and Interfaces'. * Expand the 'Chassis' tree and refresh the list by pressing F5. * If it's still not listed, power off the system, ensure the hardware is correctly installed, and restart. * Check the Device Manager to verify the PXIe-446x appears there under an NI entry. If there are error conditions, reinstall the NI-DAQmx driver.