CALIBRATION PROCEDURE
NI PXIe-4357
This document contains information about calibrating National Instruments PXIe-4357
SC Express modules using NI-DAQmx. For more information about calibration, visit ni.com/
calibration
.
Contents
Software Requirements............................................................................................................. 1
Documentation Requirements .................................................................................................. 2
Calibration Interval................................................................................................................... 2
Test Equipment......................................................................................................................... 2
Test Conditions......................................................................................................................... 3
Calibration Procedure ............................................................................................................... 3
Initial Setup....................................................................................................................... 3
Connecting the Calibrator to the Module ................................................................. 4
Verification ....................................................................................................................... 5
4-Wire Accuracy Verification .................................................................................. 5
Adjustment Procedure ...................................................................................................... 7
Specifications............................................................................................................................ 9
Where to Go for Support .......................................................................................................... 10
Software Requirements
Install NI-DAQmx 9.5 or later on the calibration computer. NI-DAQmx includes high-level
function calls to simplify the task of writing software to calibrate modules. You must have the
proper module driver installed on the calibration system before calibrating the module.
Note NI recommends that you install the NI-DAQmx driver software before
physically installing the NI PXIe-4357. NI-DAQmx, available at
ni.com/
downloads
, configures and controls the NI PXIe-4357.
NI-DAQmx supports a number of programming languages, including LabVIEW,
LabWindows
/CVI
, C/C++, C#, and Visual Basic .NET.
ni.com/manuals
DeutschFrançais

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the National Instruments PXIe-4357 and is the answer not in the manual?

National Instruments PXIe-4357 Specifications

General IconGeneral
TypeThermocouple and RTD Measurement Module
Resolution24-bit
Sample RateUp to 10 S/s per channel (aggregate)
ConnectivityPXI Express
Bus InterfacePXI Express
Operating Temperature Range0 °C to 55 °C
Storage Temperature Range-20 °C to 70 °C
Input TypeThermocouple, RTD
Thermocouple TypesB, E, J, K, N, R, S, T
RTD TypesPT100, PT1000
Resistance Measurement Range0 Ω to 400 kΩ
Voltage Measurement Range-10 V to 10 V
Accuracy±0.5 °C
Power Requirements+3.3 V
Connector TypeVHDCI
Dimensions3U
Weight300 g
Channels32 AI (High-Density)
Measurement Range (Temperature)-200 °C to 1820 °C (depending on thermocouple type)