EasyManuals Logo

National Instruments PXIe-4468 User Manual

National Instruments PXIe-4468
40 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #37 background imageLoading...
Page #37 background image
PXIe-4468 Calibration Procedure
© National Instruments Corporation 35
Call DAQmxInitExtCal (DAQmx Initialize External Calibration VI) to
initialize the calibration using the following parameters:
o deviceName: Dev_name
o password: NI
o calHandle: &myCalHandle
2. Connect the calibrator to AI0 as shown in Figure 1 or Figure 2.
3. Configure the calibrator to generate an output with the amplitude of
6.3 Vrms and frequency of 1kHz.
4. Call DAQmxAdjust4468Cal (DAQmx Adjust 4468 Calibration VI from the
DAQmx Adjust DSA AI Calibration Polymorphic AI), using the following
parameters:
calHandle: myCalHandle
referenceVoltage: 6.3
Timebase Adjustment
Perform an adjustment at least once within the calibration interval.
Adjustment automatically updates the calibration constants, the date, and
the temperature in the DUT EEPROM. If the DUT passes the verification
procedures within the As-Left test limits, an adjustment is not required.
Proceed to the Update the Onboard Calibration Information section.
Note
Initial test connection is same as Timebase Frequency Verification, as
shown in Figure 9 or Figure 10.
Adjustment Procedure
1. Connect channel AI0 to the function generator as shown in Figure 9 or
Figure 10.

Table of Contents

Other manuals for National Instruments PXIe-4468

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the National Instruments PXIe-4468 and is the answer not in the manual?

National Instruments PXIe-4468 Specifications

General IconGeneral
BrandNational Instruments
ModelPXIe-4468
CategoryControl Unit
LanguageEnglish

Related product manuals