PXIe-4468 Calibration Procedure
30 ni.com/services
Figure 8. Initial Connection for AO Common-Mode Gain and Offset-mXLR
1.
DMM
4.
BNC
mXLR Cable
2.
Banana
BNC Adapter
5.
PXIe
-4468 with mXLR
3
Shorting Cable (Banana
Banana)
Verification Procedure
Complete the following procedure to verify common-mode gain and offset.
Repeat 2 times, once for each channel.
10. Connect the AO channel under test to the DMM as shown in Figure 7 or
Figure 8.
Repeat 6 times, once for each test point.
11. Configure the DMM using the information in Table 12.
Table 24: DMM Configuration
Configuration Value
Function DC Volts (DCV)
Range Fixed - 10 V
Number of Power Line Cycles (NPLC) 100