installations of Measurement Categories CAT II,
CAT III, or CAT IV.
Caution Do not connect the NI 9217 to signals or use
for measurements within Measurement Categories III
or IV.
Safety Guidelines for Hazardous Voltages
If hazardous voltages are connected to the device, take the
following precautions. A hazardous voltage is a voltage greater
than 42.4 Vpk voltage or 60 VDC to earth ground.
Caution Ensure that hazardous voltage wiring is
performed only by qualified personnel adhering to local
electrical standards.
Caution Do not mix hazardous voltage circuits and
human-accessible circuits on the same module.
Caution Ensure that devices and circuits connected to
the module are properly insulated from human contact.
Caution When module terminals are hazardous
voltage LIVE (>42.4 Vpk/60 VDC), you must ensure
that devices and circuits connected to the module are
NI 9217 Getting Started Guide | © National Instruments | 5