EasyManua.ls Logo

National Instruments NI USB-9234 - Power Requirements; Bus Interface; Physical Characteristics; Safety Voltages

National Instruments NI USB-9234
23 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
NI USB-9234 User Guide and Specifications 20 ni.com
Power Requirements
Current consumption from USB.............500 mA, max
Suspend mode..................................2.5 mA, max
Bus Interface
USB specification ...................................USB 2.0 Hi-Speed
Physical Characteristics
Dimensions .............................................12.1 cm × 8.6 cm × 2.5 cm
(4.75 in. × 3.37 in. × 0.99 in.)
Weight ....................................................Approx. 250 g (8.8 oz)
Safety
If you need to clean the module, wipe it with a dry towel.
Safety Voltages
Connect only voltages that are within the following limits.
Channel-to-earth ground.........................±30 V max,
Measurement Category I
Isolation
Channel-to-channel..........................No isolation between channels
Channel-to-earth ground..................No isolation between channels
and earth ground
Measurement Category I is for measurements performed on circuits not
directly connected to the electrical distribution system referred to as
MAINS voltage. MAINS is a hazardous live electrical supply system that
powers equipment. This category is for measurements of voltages from
specially protected secondary circuits. Such voltage measurements include
signal levels, special equipment, limited-energy parts of equipment,
circuits powered by regulated low-voltage sources, and electronics.
Caution Do not connect the NI USB-9234 to signals or use for measurements within
Measurement Categories II, III, or IV.

Related product manuals