14 | ni.com | NI PXI-4130 Calibration Procedure
1
2 mA 20 V 20 V 1 40.0 μA ±1.01 μA
2 0.50 mA ±1.15 μA
3 1.00 mA ±1.30 μA
4 1.50 mA ±1.45 μA
5 2.00 mA ±1.60 μA
-20 V 6 40.0 μA ±1.01 μA
7 0.50 mA ±1.15 μA
8 1.00 mA ±1.30 μA
9 1.50 mA ±1.45 μA
10 2.00 mA ±1.60 μA
20 mA 20 V 1 0.40 mA ±10.1 μA
2 5.00 mA ±11.5 μA
3 10.0 mA ±13.0 μA
4 15.0 mA ±14.5 μA
1 20 mA 20 V 20 V 5 20.0 mA ±16.0 μA
-20 V 6 0.40 mA ±10.1 μA
7 5.00 mA ±11.5 μA
8 10.0 mA ±13.0 μA
9 15.0 mA ±14.5 μA
10 20.0 mA ±16.0 μA
Table 7. NI 4130 Output Parameters and Test Limits for Current Programming
Accuracy Verification (Continued)
Channel(s)
Current
Limit
Range
Voltage
Level
Range
Vol tag e
Level
Iteration
Output
(A)
Test Limit