18 | ni.com | NI PXI-4130 Calibration Procedure
1
20 mA 20 0 1 5.00 mA 0.03% + 2.0 μA
20 2 5.00 mA
3 10.0 mA
4 15.0 mA
5 20.0 mA
-20 6 5.00 mA
1 20 mA 20 V -20 7 10.0 mA 0.03% + 2.0 μA
8 15.0 mA
9 20.0 mA
200 mA 0 1 50.0 mA 0.03% + 40 μA
20 2 50.0 mA
3 100 mA
4 150 mA
5 200 mA
-20 6 50.0 mA
7 100 mA
8 150 mA
9 200 mA
Table 8. NI 4130 Output Parameters and Test Limits for Current Measurement