NI PXI-4130 Calibration Procedure | © National Instruments | 43
Table 21. NI 4130 Output Parameters and Test Limits for Current Measurement
Accuracy Adjustment
Channel
Current
Limit
Range
Voltag e
Level
Range
Voltag e
Level
Iteration Output
0 1A 6V 0V 1 350 mA
6V 2 350 mA
3 700 mA
1 200 μA 20 V 0V 1 200 µA
20 V 2
-20 V 3
2mA 0V 1 2 mA
20 V 2
-20 V 3
20 mA 0V 1 20 mA
20 V 2
-20 V 3
200 mA 0V 1 200 mA
20 V 2
-20 V 3
2A 0V 1 800 mA
20 V 2
-20 V 3