ALGO® 5 Newborn Hearing Screener Service Manual
P/N 000995 Rev.A
15
FIGURE 21: DSP Box to the mounting bracket
After the DSP Box is installed, tighten the grounding screw (10x32) as shown in
FIGURE 22 until the screw breaks the outer coating of the DSP box to come into
contact with the aluminum. This will provide an electrical connection between the
DSP mounting bracket and the DSP Box.
FIGURE 22: Grounding screw DSP Box
From the back of the cart column, identify the DSP power cord coming from the top
power supply mounted in the cart column (see FIGURE 23). Route this cable
through the top opening in the cart column, down through the cable guide on the front
of the column, and plug it into the DSP Box.
Grounding
screw