PERITONE Instruction Manual - GB
20
Specifications
1. Single Channel
2. EMG Range: 0.2 to 2000 µV RMS (continuous)
3. Sensitivity: 0.l µV RMS
4. Accuracy: 4% of µV reading +/- 0.3 µV at 200 Hz
5. Selectable Bandpass filter - 3db Bandwidth;
a. Wide: 18 Hz +/- 4 Hz to 370 Hz +/- 10% - Reading
below 235 microvolts. 10 Hz +/- 3 Hz to 370 Hz +/-
10% - Reading above 235 microvolts.
b. Narrow: 100 Hz +/- 5% to 370 Hz +/- 10%
6. Notch filter: 50 Hz (Canada 60Hz) - 33 dbs (0.1% accuracy)
7. Common Mode Rejection Ratio: 130 dbs Minimum @ 50 Hz
8. Battery: PP3 Alkaline
9. Work / Rest periods: 2-99 seconds
10. Number of Trials: 1-99
11. Low battery indication at 7.4 volts +/- 0.2 volts and
auto shut off 10 minutes after last key pressed
12. Environmental conditions for storage & transport:
-10 to +50ºC. 0-90% Humidity
13. Physical Dimensions: Length 128.5mm, Width 64mm,
Depth 28.3 mm
14. Weight: 0.15 Kg