28
Emissions test Compliance Electromagnetic environment –
guidance
RF emissions
CISP R 11
Group 1The PERITONE
product uses RF
energy only for its internal function.
Therefore, its RF emissions are very
low and are not likely to cause any
interference in nearby electronic
equipment.
Table 201: Guidance and manufacturer’s declaration
– electromagnetic emissions
product
The PERITONE product is intended for use in the electromagnetic
environment specified below. The customer or the user of the PERITONE
should ensure that it is used in such an environment.
RF emissions
CISP R 11
Class B
Harmonic emissions
IEC 61000-3-2
IEC 61000-3-2
Not applicable
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Not applicable
The PERITONE product is suitable
for use in all establishments,
including domestic establishments
and those directly connected to the
public low-voltage power supply
network that supplies buildings used
for domestic purposes.
Immunity test IEC 60601
test level
Compliance
level
Electromagnetic environment
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wo od, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative
humidity should be at least 30 %.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic fields should
be at characteristic levels of a typical
location in a typical commercial or
hospital environment.
Table 202: Guidance and manufacturers declaration
– electromagnetic immunity
The PERITONE product is intended for use in the electromagnetic environment specified
below. The cust omer or the user of the PERITONE product should ensure that it is used in